Digital multimeter repair tips

For a faulty instrument, first check and judge the fault phenomenon is common (all functions can not be measured), or personality (individual function or individual range), and then distinguish the situation, symptomatic solution.

1. If all the files are not working, the power supply circuit and the A/D converter circuit should be checked. When checking the power supply part, you can remove the laminated battery, press the power switch, connect the meter with the positive meter to the negative, the negative meter is connected to the power supply (for the digital multimeter), and the switch hits the diode measurement file. The forward voltage of the diode indicates that the power supply is good. If the deviation is large, there is a problem with the power supply. If an open circuit occurs, focus on the power switch and battery leads. If a short circuit occurs, you need to use the open circuit method to gradually disconnect the components that use the power supply, and focus on checking the operational amplifier, timer, and A/D converter. In the event of a short circuit, it is generally not necessary to damage an integrated component. Check that the A/D converter can be performed simultaneously with the basic table, which is equivalent to the DC header of the analog multimeter. The specific inspection method is as follows:

(1) The range of the meter to be tested is transferred to the lowest level of the DC voltage;

(2) Measure whether the working voltage of the A/D converter is normal. According to the type of A/D converter used in the table, corresponding to the V+ and COM feet, the measured value is compared with its typical value.

(3) Measure the reference voltage of the A/D converter. The reference voltage of the commonly used digital multimeter is generally 100mV or 1V, that is, measure the DC voltage between VREF+ and COM. If it deviates from 100mV or 1V, it can pass the external potentiometer. Make adjustments.

(4) Check the number of inputs with zero input, short the positive terminal IN+ of the A/D converter to the negative terminal IN-, so that the input voltage Vin=0, the meter displays “00.0” or “00.00”.

(5) Check the full bright stroke of the display. The test terminal TEST pin is shorted to the positive power supply terminal V+ to make the logic ground high, and all digital circuits stop working. Since DC voltage is applied to each stroke, all the strokes on the alignment table display "1888" and the alignment table displays "18888". If there is a lack of strokes, check the A/D converter's corresponding output pin and conductive paste (or wire), and whether there is contact and disconnection between the display and the display.

2. If there is a problem with the individual files, the A/D converter and the power supply unit are working properly. Because DC voltage and resistance files share a set of voltage dividing resistors; AC and DC currents share a shunt; AC voltage and AC current share a set of AC/DC converters; others such as Cx, HFE, F, etc. are composed of independent converters. . Knowing the relationship between them, and then according to the power map, it is easy to find the fault location. If the measurement small signal is inaccurate or the display digital jitter is large, it is important to check whether the contact of the range switch is good.

3. If the measured data is unstable and the value always increases cumulatively, short-circuit the input of the A/D converter and the display data is not zero, which is generally caused by poor performance of the reference capacitor of 0.1μF.

According to the above analysis, the basic sequence of repair of the digital multimeter should be: digital meter head → DC voltage → DC current → AC voltage → AC current → Resistance file (including buzzer and check diode positive pressure drop) → Cx → HFE , F, H, T, etc. But you can't overdo it. Some problems that can be clearly seen can be dealt with first. However, when making adjustments, you must follow the above procedure.

In short, a fault multimeter, after proper detection, first analyzes the parts where the fault may occur, and then finds the fault location according to the line map for replacement and repair. Because the digital multimeter is a relatively sophisticated instrument, the components must be replaced with the same parameters, especially the replacement of the A/D converter. It must be strictly integrated by the manufacturer. Otherwise, errors will occur and the required Accuracy. The newly changed A/D converter also needs to be checked according to the method described above, and must not be trusted because of the new.

At present, there are many manufacturers of digital multimeters in China, and the quality is also good or bad. The quality of double-sided copper plates is not easy to find in repairs. When the insulation strength of the resin board is insufficient, the error is mainly caused when the high voltage is measured, and the resistance value of the voltage dividing resistor is distinguished from the change when the repair is performed. In this case, it is best to use the open circuit method to find the point of failure. The burnt carbonized parts should be cleaned to meet the insulation requirements. When the double-sided connection is broken due to the breakage of the transition hole, it is easy to be confused with the poor transfer switch and it is difficult to separate. This type of fault should be short-circuited to find the fault point. (end)

Pocket Folders

Keep your documents organized and protected with these Plastic Presentation Folders. They are ideal for school, office, home and much more. They have a smooth finish These folders measure 9" x 12" and feature assorted colors.

Made with highly durable, flexible, smooth plastic and includes business card slot on right side pocket.

6337

Multi Pocket Folder,Pocket Folders,Custom Folders,Custom Printed Pocket Folders

shaoxing chaofeng stationery manufacturing CO.,LTD. , https://www.chaofengstationery.com